12-7
Troubleshooting
Operator Manual
129373-496
Screen with Service Instructions
Screen with Operator Instructions
Occurs if steam
pressure in door seal
drops below 5 psig
(0.34 Pbar).
12.2.8 Chamber
Pressure/
Temperature
Failure
Occurs if chamber
pressure or tempera-
ture readings are
outside the normal
steam range during
sterilize phase.
12.2.7 Door Unsealed
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
207
DOOR UNSEALED
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
•
EXHAUST OR AIR BREAK CHAMBER TO
ATMOSPHERIC PRESSURE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
→
→
→
→
→
STEAM PRESSURE IN DOOR SEAL BELOW
5 psig
CAUSES AND CORRECTION:
1.
SEAL PRESSURE SWITCH MALFUNCTION
→
→
→
→
→
CHECK PS1(PS2) CONNECTIONS
→
→
→
→
→
READJUST PS1(PS2)
→
→
→
→
→
REPAIR PS1(PS2)
2.
SEAL NOT ACTIVATED
→
→
→
→
→
CHECK SEAL STEAM
→
→
→
→
→
CHECK SEAL EXHAUST
3.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S35(S36)
STATUS..
SERVICE INFORMATION:
208
DOOR UNSEALED
PAPER
CLEAR
EXIT
FEED
ALARM
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM! CHAMBER
219
PRESSURE/TEMPERATURE FAILURE
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
Alarm
Description
Screen with Operator Instructions