8-14
129373-496
Operating Instructions
Programming Cycle Values
#3 = Door B is designated as the operating end (OE). Sterilizer must be
loaded from door B, and can be unloaded from door A or door B.
At end of cycle, operator must manually unseal the unload door by
pressing
UNSEAL DOOR
touch screen button (screen #65)
located on the same end as the door.
NOTE: Once the operating end (OE) door is opened, the non-
operating end (NOE) door cannot be opened until a complete
cycle is run. If the NOE door is opened, the OE door cannot be
opened until the NOE door is closed and sealed.
#4 = Door B is designated as the operating end (OE). Sterilizer must be
loaded from door B and unloaded from door A. At end of cycle,
door A is automatically unsealed.
NOTE: Once the operating end (OE) door is opened, the non-
operating end (NOE) door cannot be opened until a complete
cycle is run. If the NOE door is opened, the OE door cannot be
opened until the NOE door is closed and sealed.
#5 = Door A is designated as the operating end (OE) door. Sterilizer
must be loaded and unloaded from door A; door B is locked out.
At end of cycle, door A is automatically unsealed.
#6 = Door B is designated as the operating end (OE) door. Sterilizer
must be loaded and unloaded from door B; door A is locked out.
At end of cycle, door B is automatically unsealed.
7. After all cycle value changes have been made, press
CYCLE MENU
button
on screen #22. Display returns to screen #15.
B (OE)
A (NOE)
➤
➤
➤
➤
➤
➤ ➤ ➤ ➤ ➤
load
unload
B (OE)
A (NOE)
➤
➤
➤
➤
➤
➤
➤
➤
➤
➤
load
unload
B (OE)
A (NOE)
➤
➤
➤
➤
➤
➤
➤
➤
➤
➤
load
unload
load
unload
B (OE)
➤
➤
➤
➤
➤
➤ ➤ ➤ ➤ ➤
A
locked
out
➤ ➤ ➤ ➤ ➤
➤
➤
➤
➤
➤
load
unload
A (OE)
B
locked
out
Pressing PREVIOUS returns display to
screen #21.
Pressing MAIN MENU returns display to
screen #1.
22
CHANGE VALUES - 1. PREVAC (PREVAC)
PRINT INT = 2 min
VACUUM DRY = 10.0 inHg
DRY TIME = 5:00
INTERLOCK = 2
PREVIOUS
MAIN
MENU
CYCLE
MENU
Double Door
Units Only