9-4
129373-496
Operating Instructions
Programming Operating Parameters
Sterilizer uses the "too long in phase" values to monitor and control the length
of the associated cycle phases. Each "too long in phase" value is factory
programmed at 30 minutes.
To adjust the "too long in phase" values:
1. Access Change Values menu (screen #14) as described at the beginning
of this section.
2. Press
TOO LONG IN PHASE
.
9.3 Too Long In
Phase
14
CHANGE VALUES
RETURN
MAIN
MENU
TOO LONG
IN PHASE
TIME/
DATE
Pressing RETURN returns display to
screen #13.
Pressing MAIN MENU returns display to
screen #1.
3. Screen #40 is the first of two menu screens which list the phase and its
currently set value. To change any of the listed values (parameters), press
the corresponding button.
NOTE: A definition of each value (parameter) is given on the corresponding
display screen.
Pressing RETURN saves all changed
parameters and returns display to screen
#14.
Pressing NEXT advances display to
screen #53.
Pressing MAIN MENU returns display to
screen #1.
40
CHANGE TOO LONG IN PHASE PARAMETERS
TOO LONG IN:
CHARGE = 30:00
STERILIZE = 30:00
EXHAUST = 30:00
EVACUATE = 30:00
NEXT
RETURN
MAIN
MENU
CYCLES
SETUP