8-13
Programming Cycle Values
Operating Instructions
129373-496
•
If sterilizer is equipped with double doors, pressing
INTERLOCK
ad-
vances display to screen #8. Enter interlock type using the numeric
keypad. Interlock type appears on the display as it is being entered.
NOTE: Default interlock type is #2. If different setting is desired, a
separate interlock type must be set for each cycle.
8
CHANGE VALUES - INTERLOCK TYPE
RETURN
1
2
3
4
5
6
<-
0
->
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
cycle value and returns display to screen
#22.
SET INTERLOCK TYPE
2
The interlock type
determines which door
is used to load and unload the
sterilizer.
See operating manual for
description of types.
The following descriptions and illustrations explain each interlock type,
#0 through #6. Each door on the illustrations is labeled, "A" or "B", for
reference. Door A is located on the same end as the main power
disconnect switch (see Figure 3-2 for switch location). Door B is located
on the opposite end.
#0 = No interlocks. Either door can be used to load and unload the
sterilizer. Both doors can be open at the same time. At end of
cycle, both doors are automatically unsealed.
#1 = Door A is designated as the operating end (OE). Sterilizer must be
loaded from door A, and can be unloaded from door A or door B.
At end of cycle, operator must manually unseal the unload door by
pressing the
UNSEAL DOOR
touch screen button (screen #65)
located on the same end as the door.
NOTE: Once the operating end (OE) door is opened, the non-
operating end (NOE) door cannot be opened until a complete
cycle is run. If the NOE door is opened, the OE door cannot be
opened until the NOE door is closed and sealed.
#2 = Door A is designated as the operating end (OE). Sterilizer must be
loaded from door A and unloaded from door B. At end of cycle,
door B is automatically unsealed.
NOTE: Once the operating end (OE) door is opened, the non-
operating end (NOE) door cannot be opened until a complete
cycle is run. If the NOE door is opened, the OE door cannot be
opened until the NOE door is closed and sealed.
B (NOE)
A (OE)
➤ ➤ ➤ ➤ ➤
➤ ➤ ➤ ➤ ➤
load
unload
➤ ➤ ➤ ➤ ➤
➤
➤
➤
➤
➤
load
unload
B (NOE)
A (OE)
B (NOE)
A (OE)
➤ ➤ ➤ ➤ ➤
➤ ➤ ➤ ➤ ➤
load
unload