STERILIZER OPERATION
6.1 Before
Operating Sterilizer
The following steps must be performed prior to daily sterilizer usage.
1. Open chamber door and check that drain strainer is clean and in place.
2. Check that chamber interior is clean and close chamber door. Refer to
S
ECTION
11, R
OUTINE
M
AINTENANCE
, later in this manual, if cleaning is necessary.
3. Open front cabinet panel on load end of the sterilizer. Verify that steam and
water supply valves to the sterilizer are on (see Figure 6-1). Close cabinet
panel.
4. Open printer access door. Check that sufficient amount of printer paper is
available. A colored warning stripe is visible when paper roll is near the end.
Refer to
S
ECTION
11, R
OUTINE
M
AINTENANCE
, if the paper roll needs replaced.
6
WARNING – SLIPPING HAZ-
ARD: To prevent falls, keep
floors dry by immediately wip-
ing up any spilled liquids or
condensation in sterilizer
loading and unloading areas.
Main Power
Disconnect Switch
Chamber Exhaust Valve
(Door Seal Exhaust Valve Not Shown)
Steam Supply Valve
Water Supply
Valve
Figure 6-1. Steam and Water Supply Valves
NOTE: Valves and dis-
connect switch enclosed
within cabinet, or behind
wall enclosure (depend-
ing on configuration).