6-21
Sterilizer Operation
Operating Instructions
129373-496
• Charge
– Chamber charges with steam to the programmed sterilize
temperature.
• Sterilize
– Sterilize phase begins when chamber temperature is equal to
or greater than the programmed sterilize temperature. Chamber tempera-
ture is printed every 2 minutes (or printed after each programmed print
interval). The steam-to-jacket valve is regulated to maintain the chamber
at the sterilize control temperature (control temperature = sterilize tem-
pe overdrive temperature).
9
LIQUID CYCLE
PHASE:
CHARGE
CHARGING TO:
121.0 C
CHAMBER:
110.0 C
10.0
psig
PROJECTED CYCLE COMPLETION TIME:
58:40
m:s
STATUS
ABORT
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
9
LIQUID CYCLE
PHASE:
STERILIZE
TIME LEFT IN PHASE:
30:00
CHAMBER:
121.0 C
19.0
psig
PROJECTED CYCLE COMPLETION TIME:
50:40
m:s
STATUS
ABORT
• Slow Exhaust
– Chamber is slowly exhausted until chamber pressure
reaches 0 psig (0 bar).
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
9
LIQUID CYCLE
PHASE:
SLOW EXHAUST
EXHAUSTING TO:
0.0 psig
CHAMBER:
122.5 C
19.0
psig
PROJECTED CYCLE COMPLETION TIME:
20:40
m:s
STATUS
ABORT
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.