
6-7
Sterilizer Operation
Operating Instructions
129373-496
• Activate Seal
– Door seal fills with steam and expands against the
sterilizer door opening, forming an airtight seal.
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
• Purge
– Steam flows through the chamber for the programmed time
interval.
9
PREVAC CYCLE
PHASE:
PURGE
TIME LEFT IN PHASE:
1:00
CHAMBER:
32.0 C
00.0 psig
PROJECTED CYCLE COMPLETION TIME:
32:30
m:s
ABORT
STATUS
NOTE: The projected cycle completion time shown on the display is
estimated. The control automatically evaluates the cycle progress and
corrects the estimated time at the beginning of each phase.
NOTE: Current time, chamber pressure and chamber temperature are printed
at each transition point.
9
PREVAC CYCLE
PHASE:
ACTIVATE SEAL
DOOR IS WAITING TO BE SEALED
CHAMBER:
32.0 C
00.0 psig
PROJECTED CYCLE COMPLETION TIME:
32:50
m:s
ABORT
STATUS