5-3
Control Interface
Operating Instructions
129373-496
Pressing a
cycle button
advances display to a screen listing the correspond-
ing cycle parameters.
For example: If PREVAC button on screen #2 is pressed, the cycle parameters
screen #4 appears. Screen lists the cycle parameters programmed for the
selected prevac cycle. Similar cycle parameters screens appear after press-
ing GRAVITY button and LIQUID button.
NOTE: Processing cycle parameters can be changed by the operator/super-
visor. Refer to S
ECTION
8, P
ROGRAMMING
C
YCLE
V
ALUES
.
4
CYCLE PARAMETERS
PREVIOUS
START
CYCLE
Pressing PREVIOUS returns display to
screen #2.
Pressing
START CYCLE
initiates the selected cycle and advances display to
the first in-cycle status screen (#9). Refer to
S
ECTION
6, S
TERILIZER
O
PERATION
,
before running a processing cycle.
1. PREVAC (PREVAC)
PURGE TIME ......................................... 1:00
PULSES .................................................. 4
MAX: 26.0 psig MIN: 10.0
inHg
STERILIZE:
TEMP .................................................. 132.0 C
TIME ................................................... 0:04:00
VACUUM DRY ....................................... 10.0 inHg
DRY TIME ............................................... 0:05:00
Cycle Name
(Cycle Type)