6-10
129373-496
Operating Instructions
Sterilizer Operation
• Vacuum Dry
– A vacuum is drawn in the chamber to the programmed
vacuum dry point.
9
PREVAC CYCLE
PHASE:
VACUUM DRY
EVACUATING TO:
10.0 inHg
CHAMBER:
83.5 C
2.0 psig
PROJECTED CYCLE COMPLETION TIME:
06:10
m:s
STATUS
ABORT
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
NOTE: If dry time is programmed for 0 minutes, cycle will automatically skip
the Vacuum Dry, Dry and Air Break phases; retract door seal; and remove
vapors for 1 minute before completing cycle.
• Dry
– Dry phase begins once vacuum level in chamber reaches the
programmed vacuum dry point. Chamber then continues to evacuate for
the programmed time interval.
9
PREVAC CYCLE
PHASE:
DRY
TIME LEFT IN PHASE:
5:00
CHAMBER:
83.5 C
10.0 inHg
PROJECTED CYCLE COMPLETION TIME:
05:30
m:s
STATUS
ABORT
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
• Air Break
– Filtered air enters the chamber to relieve the vacuum within the
chamber.
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
9
PREVAC CYCLE
PHASE:
AIR BREAK
AIR BREAK TO:
2.0 inHg
CHAMBER:
83.5 C
25.0 inHg
PROJECTED CYCLE COMPLETION TIME:
00:30
m:s
STATUS
ABORT