8-1
Programming Cycle Values
Operating Instructions
129373-496
PROGRAMMING CYCLE VALUES
The Amsco Century Medium Steam Sterilizer is factory programmed with
default processing cycles and cycle values (see Table 8-1). These preset
values (parameters) can be modified to process varying loads which occur in
the sterilizer's operating environment.
NOTE: Control will default to the factory-programmed values if a battery or
battery-powered memory failure occurs.
This section outlines how to change the cycle values only. For details on
modifying the sterilizer operating parameters, refer to
S
ECTION
9, P
ROGRAMMING
O
PERATING
P
ARAMETERS
.
Table 8-1. Default Processing Cycles and Cycle Values (Parameters)
Prevacuum Cycle (Cycles 1 and 4)
for sterilizing heat- and moisture-
stabile materials utilizing vacuum-
assisted air removal process.
Purge Time
=
1:00
Pulses
=
4
Max. Pressure =
26.0 psig
Min. Pressure
=
10.0 inHg
Sterilize Time
=
4:00
Sterilize Temp. =
132.0°C
Overdrive
=
1.5°C
Under Temp.
=
1.0°C
Over Temp.
=
6.0°C
Print Interval
=
2 min
Vacuum Dry
=
10.0 inHg
Dry Time
=
5:00
Interlock*
=
2
Gravity Cycle (Cycles 2 and 5)
for sterilizing heat- and moisture-
stabile materials.
Purge Time
= 1:00
Sterilize Time
= 15:00
Sterilize Temp. = 132°C
Overdrive
= 1.5°C
Under Temp.
= 1.0°C
Over Temp.
= 6.0°C
Print Interval
= 2 min
Vacuum Dry
= 10.0 inHg
Dry Time
= 5:00
Interlock*
= 2
Liquid Cycle (Cycles 3 and 6)
for sterilizing liquids and media in
vented borosilicate glass or metal
containers.
Purge Time
= 1:00
Sterilize Time
= 45:00
Sterilize Temp. = 121°C
Overdrive
= 1.5°C
Under Temp.
= 1.0°C
Over Temp.
= 6.0°C
Print Interval
= 2 min
Interlock*
= 2
* Double-Door Units Only
8