12-11
Troubleshooting
Operator Manual
129373-496
The following alarm screens will appear only when the sterilizer is
not
processing
a cycle.
12.3 Out-of-cycle
Alarms
Occurs if door switch
does not make contact
within allotted time.
12.3.1
Too Long
To Close Door
STATUS..
SERVICE INFORMATION:
240
TOO LONG TO CLOSE DOOR
→
→
→
→
→
DOOR SWITCH DID NOT MAKE IN
ALLOTTED TIME
CAUSES AND CORRECTION:
1.
DOOR SWITCH MALFUNCTION
→
→
→
→
→
CHECK LS1(LS2) CONNECTIONS
→
→
→
→
→
READJUST LS1(LS2)
→
→
→
→
→
REPAIR LS1(LS2)
2.
POWER DOOR MECHANISM FAILURE
→
→
→
→
→
REPAIR MECHANISM
→
→
→
→
→
REPLACE MOTOR
PAPER
CLEAR
EXIT
FEED
ALARM
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
239
TOO LONG TO CLOSE DOOR
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION
UNTIL DOOR IS CLOSED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK DOOR FOR OBSTRUCTION
→
→
→
→
→
REMOVE OBSTRUCTION AND
CLOSE DOOR
3. IF DOOR WILL NOT CLOSE, CALL
SERVICE
Screen with Service Instructions
Alarm
Description
Screen with Operator Instructions