11-4
129373-496
Operator Manual
Routine Maintenance
11.2 Weekly
Maintenance
11.2.1 Flush Chamber
Drain
3. Once the chamber is clean, replace the shelf assembly using the reverse of
the appropriate procedure given in step 1. Close the chamber emergency
manual exhaust valve.
4. Professional cleaning of the chamber on a yearly basis (or as required due to
local conditions) is suggested to maintain appearance of the chamber interior.
Contact STERIS for information regarding this service.
Flush chamber drain as follows whenever the line becomes clogged:
1. Turn off steam supply valve. Wait until jacket pressure is zero. Wait until
chamber has cooled to room temperature.
2. Remove chamber drain strainer (Figure 11-1). Clean strainer using
procedures given above, if necessary.
3. Pour a solution of 60 mL (approximately 1/4 cup) of STERIS Sonic
Detergent* or LiquiJet* and 500 mL (approximately1 pint) of hot water into
the drain. Solution may puddle in the bottom of the chamber.
OR
4. Should these detergents be unavailable, you may use a hot solution of 15
mL (approximately 1 tablespoon) of tri-sodium phosphate to 500 mL
(approximately1 pint) of hot water.
5. Open door and place strainer back in drain.
* Contact your local STERIS repre-
sentative.
CAUTION: Never use a wire
brush, abrasives, or steel
wool on door and chamber
assembly. Do not use clean-
ers containing chloride on
stainless-steel surfaces.
Chloride-based cleaners
will deteriorate stainless
steel, eventually leading to
failure of the vessel.
CAUTION: Lifting the cham-
ber float switch when clean-
ing the chamber may cause
the sterilizer control to ini-
tiate a "Chamber Flooded"
alarm. If this alarm condi-
tion occurs, the operator
must turn the control power
OFF then ON to clear the
alarm. The control power
switch is located in the me-
chanical area at the side of
the sterilizer. Placing the
sterilizer in standby does
not clear this alarm.
W A R N I N G – B U R N
HAZARD: Allow sterilizer
and accessories to cool
t o r o o m t e m p e r a t u r e
before performing any
cleaning or maintenance
procedures.