6-33
Sterilizer Operation
Operating Instructions
129373-496
WARNING – BURN HAZARD:
Sterilizer, rack/shelves, and
loading car will be hot after
cycle is run. Always wear pro-
tective gloves and apron (also
a face shield, if processing liq-
uids) when removing a pro-
cessed load. Protective gloves
and apron must be worn when
reloading sterilizer following
the previous operation.
WARNING – SLIPPING HAZ-
ARD: To prevent falls, keep
floors dry by immediately wip-
ing up any spilled liquids or
condensation in sterilizer load-
ing and unloading areas.
2. Screen #11 allows operator a final chance to continue with the current cycle
in progress instead of aborting cycle operation.
3. Press
ABORT
on screen #11. Printer records time the cycle was aborted.
• If cycle is aborted while pressure is in the chamber, cycle automatically
advances to the exhaust phase. Control system safely exhausts chamber
and removes vapors before unsealing the chamber door.
• If cycle is aborted while a vacuum is in the chamber, cycle automatically
advances to the air break phase. Control system safely relieves chamber
vacuum before unsealing the chamber door.
4. Once door is unsealed, display returns to the main menu (screen #1) and
sterilizer may be unloaded.
IMPORTANT:
The operator/supervisor must decide if the chamber load must
be reprocessed after the cycle was aborted.