9-5
Programming Operating Parameters
Operating Instructions
129373-496
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
parameter and returns display to screen
#40.
•
Pressing
CHARGE
advances display to screen #41. Enter charge phase
value using the numeric keypad. Value appears on display as it is being
entered.
41
CHANGE VALUES - TOO LONG IN CHARGE
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET TOO LONG IN CHARGE
30:00
(Minutes:Seconds)
The too long in
charge is the
allowable amount of
time that the chamber
has to reach the set
point during charge
pulse or charge phase
before an alarm occurs.
•
Pressing
STERILIZE
advances display to screen #52. Enter sterilize
phase value using the numeric keypad. Value appears on display as it is
being entered.
•
Pressing
EXHAUST
advances display to screen #42. Enter exhaust
phase value using the numeric keypad. Value appears on display as it is
being entered.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
parameter and returns display to screen
#40.
42
CHANGE VALUES - TOO LONG IN EXHAUST
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET TOO LONG IN EXHAUST
30:00
(Minutes:Seconds)
The too long in
exhaust is the
allowable amount of
time that the chamber
has to reach 4 psig
before an alarm occurs.
52
CHANGE VALUES - TOO LONG IN STERILIZE
RETURN
1
2
3
4
5
6
7
8
9
←
←
←
←
←
0
→
→
→
→
→
SET TOO LONG IN
STERILIZE
30:00
(Minutes:Seconds)
The too long in
sterilize is the
allowable amount of
time that the sterilize
phase may go beyond the set
sterilize time. The
cycle will then abort.
Pressing
←
or
→
on numeric keypad
moves the cursor to the left or right,
respectively.
Pressing RETURN saves the changed
parameter and returns display to screen
#40.
Definition of
Parameter