Operator Manual
Troubleshooting
Alarm
Description
Screen with Service Instructions
Screen with Service Instructions
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
289
DOOR PRESSURE FAILURE
CHAMBER:
00.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
•
EXHAUST OR AIR BREAK CHAMBER TO
ATMOSPHERIC PRESSURE
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
STATUS..
SERVICE INFORMATION:
290
DOOR PRESSURE LOCK:
→
→
→
→
→
STEAM PRESSURE IN DOOR SEAL BELOW
5 psig
CAUSES AND CORRECTION:
1.
SEAL PRESSURE SWITCH MALFUNCTION
→
→
→
→
→
CHECK LS5(LS6) CONNECTIONS
→
→
→
→
→
READJUST LS5(LS6)
→
→
→
→
→
REPAIR LS5(LS6)
2.
SEAL NOT ACTIVATED
→
→
→
→
→
CHECK SEAL STEAM
→
→
→
→
→
CHECK SEAL EXHAUST
3.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S35 (S36)
PAPER
CLEAR
EXIT
FEED
ALARM
Reference Deviation
Alarm, Continued
STATUS..
SERVICE INFORMATION:
250
REFERENCE/CHAMBER DEVIATION
→
→
→
→
→
DEVIATION BETWEEN REFERENCE AND
CHAMBER PROBE IS MORE THAN 1 C.
CAUSES AND CORRECTION:
1.
LOOSE CONNECTION IN PROBE WIRING
→
→
→
→
→
REPAIR
2.
CHAMBER REFERENCE PROBE FAILED
→
→
→
→
→
REPLACE
→
→
→
→
→
RECALIBRATE
3.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
4.
MAIN CONTROL FAILURE
→
→
→
→
→
CHECK CONTROL BOARD STATUS
LEDs
→
→
→
→
→
REPLACE 146656-185 CONTROL BOARD
→
→
→
→
→
REPLACE 146659-064 ANALOG BOARD #1
PAPER
CLEAR
EXIT
FEED
ALARM
Alarm
Description
Screen with Operator Instructions
12.2.12 Door Pressure
Lock
(Hinged Door Models
Only)
Occurs if door pressure
lock is sensed engaged
while in cycle.