6-8
129373-496
Operating Instructions
Sterilizer Operation
• Vacuum Pulse Exhaust
– Chamber is exhausted to 4 psig (0.28 bar).
NOTE: The sterilizer can be programmed to pull up to 99 pulses during the
Prevacuum cycle. Refer to S
ECTION
8, P
ROGRAMMING
C
YCLE
V
ALUES
.
9
PREVAC CYCLE
PHASE:
VACUUM PULSE #1
EXHAUSTING TO:
4.0 psig
CHAMBER:
75.0 C
10.0
psig
PROJECTED CYCLE COMPLETION TIME:
31:00
m:s
STATUS
ABORT
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
• Vacuum Pulse Evacuate
– A vacuum is drawn in the chamber to the
programmed minimum pressure parameter.
9
PREVAC CYCLE
PHASE:
VACUUM PULSE #1
EXHAUSTING TO:
10.0 inHg
CHAMBER:
75.0 C
8.0 inHg
PROJECTED CYCLE COMPLETION TIME:
29:40
m:s
STATUS
ABORT
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
Pressing ABORT advances display to
screen #11. Refer to "Aborting Cycles,"
later in this section.
Pressing STATUS PRINT generates a
printout of the current sterilizer chamber
status.
• Charge Pulse
–
Chamber charges with steam to the programmed maxi-
mum pressure parameter.
9
PREVAC CYCLE
PHASE:
CHARGE PULSE #1
STEAM CHARGING TO:
26.0 psig
CHAMBER:
65.0 C
10.0 inHg
PROJECTED CYCLE COMPLETION TIME:
29:00
m:s
STATUS
ABORT