9-1
Programming Operating Parameters
Operating Instructions
129373-496
PROGRAMMING OPERATING PARAMETERS
The Amsco
®
Century
TM
Medium Steam Sterilizer 26" x 37.5" is factory pro-
grammed with default cycle values and operating parameters. The operating
parameters are used to control the general way a sterilizer operates. Sterilizer
operating parameters include time/date, too long in phase values and setup
values.
NOTE: Control will default to the factory-programmed values if a battery or
battery-powered memory failure occurs.
This section outlines how to change the operating parameters only. For details
on modifying the default cycle values, refer to
S
ECTION
8, P
ROGRAMMING
C
YCLE
V
ALUES
.
9.1 Access Change
Values Menu
1
S T E R I S S C I E N T I F I C
STERILIZER PREPARED FOR:
STERIS SCIENTIFIC
CUSTOMER
MAIN MENU
CYCLE
SELECT
OPTIONS
2. Press
CHANGE VALUES
. Printer records the date and time when Change
Values option was selected.
NEXT
MAIN
MENU
Pressing NEXT advances display to
screen #87.
Pressing MAIN MENU returns display to
screen #1.
13
OUT OF CYCLE OPTIONS
STATUS
CHANGE
VALUES
DUPLICATE
DISPLAY
VALUES
PAPER
FEED
VALUES
NOTE: If access code feature is activated, an assigned four-digit code must be
correctly entered before operator can change parameters. Refer to S
ECTION
7,
E
NTERING
A
CCESS
C
ODE
.
To modify the preset sterilizer operating parameters, access the Change
Values menu (screen #14) as follows:
1. Press
OPTIONS
on screen #1.
Pressing CYCLE SELECT advances
display to the first Cycle Select menu
(screen #2).
9