CYCLES
RECOMMENDED
STERILIZE
STERILIZE
DRY TIME
LOADS
TEMP.
TIME
PREVACUUM
Heat and moisture-stabile
132°C
4 minutes
5 minutes
materials utilizing vacuum-
(270°F)
assisted air removal process.
GRAVITY
Heat and moisture-stabile
132°C
15 minutes
5 minutes
materials
(270°F)
LIQUIDS
Liquids and media in
121°C
45 minutes
N/A
vented borosilicate glass
(250°F)
containers.
The Amsco Century Medium Steam Sterilizer is equipped with the following default cycles and cycle
values (see Section 7 for detailed cycle parameters and instructions for modifying cycle parameters).