12-5
Troubleshooting
Operator Manual
129373-496
Screen with Operator Instructions
Occurs if chamber
temperature drops
below sterilize
temperature.
Screen with Service Instructions
Too Long In Air Break
(Continued)
NOTE: If under tem-
perature value is set to
"RESUME,") the
sterilizer will automati-
cally resume sterilize
time after set tempera-
ture is reached.
12.2.5 Under Sterilize
Temperature
Alarm
Description
Screen with Service Instructions
STATUS..
SERVICE INFORMATION:
226
TOO LONG IN AIR BREAK
→
→
→
→
→
CHAMBER DID NOT AIR BREAK VACUUM
TO 2 inHG WITHIN ALLOTTED TIME
CAUSES AND CORRECTION:
1.
AIR INLET FILTER PLUGGED
→
→
→
→
→
REPLACE
2.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S01
3.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
SERVICE
EXIT
FEED
MODE
STATUS
PAPER
FEED
SILENCE
SERVICE
ALARM
HELP
STATUS..
ALARM!
223
UNDER STERILIZE TEMPERATURE
CHAMBER:
000.0 C
0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY RESTART STERILIZE TIMER
AND CONTINUE CYCLE AFTER SET
TEMP. IS REACHED
OPERATOR INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
STATUS..
SERVICE INFORMATION:
224
UNDER STERILIZE TEMPERATURE
→
→
→
→
→
CHAMBER TEMPERATURE DROPPED BELOW
STERILIZE TEMPERATURE BY UNDERTEMP
TEMPERATURE VALUE
CAUSES AND CORRECTION:
1.
STEAM PRESSURE LESS THAN 50 PSIG
→
→
→
→
→
CHECK STEAM SUPPLY PIPING
2.
CHAMBER STEAM TRAP MALFUNCTION
→
→
→
→
→
REPAIR
3.
SOLENOID VALVE MALFUNCTION
→
→
→
→
→
REPAIR S09
→
→
→
→
→
REPAIR S02
4.
CONTROL OUT OF CALIBRATION
→
→
→
→
→
RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
SERVICE
EXIT
FEED
MODE