Asterion DC Multioutput ASA Series Operational Manual
M330516-01
4-7
Figure 4-5: Remote Inhibit Mode
Note
:
The data format <n>represents individual channel number 1, 2 and 3
.
LATC(hing)
An open contact between remote inhibit terminal and return
latches the output in the protection shutdown state; this state can
be cleared either through the front panel fault screen (refer Figure
4-7) or by issuing the SCPI command, OUTP<n>:PROT:CLE.
Figure 4-6: Remote Inhibit Clear Fault Screen
LIVE
The output state follows the state of the Remote Inhibit input. An
open contact between the remote inhibit terminal and return will
disable the output; a contact closer will clear the fault.
Factory default mode is LIVE. With this default factory setting (connected remote
inhibit 26 pin mating connector installed and the remote inhibit mode being LIVE), the
user can turn ON and OFF the power supply with the output switch.
User can remove the hardwired jumpers in the remote inhibit 26 pin female connector
and wire to a contact closure as per their requirement. User would need to select the
mode of remote inhibit using the front panel screen refer Figure 4-5, according to their
requirement of fault clearance.
For additional information on programming the Remote Inhibit function, refer to the
Asterion DC Multioutput Programming Manual P/N M330517-01. Refer to AMETEK
Programmable Power website, www.powerandtest.com, to download latest version.
correlate
4.3
Remote Output ON/OFF Control by External
Source
Application of DC voltage (2.7-24V) between Pins 4 and 8, will Enable the Output of
Channel 1, DC voltage (2.7-24V) between Pins 13 and 17, will Enable the Output of
Channel 2 and DC voltage (2.7-24V) between Pins 22 and 26, will Enable the Output
of Channel 3. This interface is opto-isolated from circuit common. See Figure 4-7,
Figure 4-8 and Figure 4-9 for connection requirements.