Asterion DC Multioutput ASA Series Operational Manual
M330516-01
3-64
Figure 3-74: PONS - Measurement Settings Screen
The Measurement Settings screen has the following fields:
Entry
Description
Volt Avg Samples
Sets the number of voltage readings to average in the voltage
readback during Unit power-on.
Allows to set a value between 1 to 5. The value of 1 (factory default)
provides the fastest response time in the readings, but less rejection
of noise.
Curr Avg Samples
Sets the number of current readings to average in the current readback
during the power-on.
Allows to set a value between 1 to 9. The value of 1 (factory default)
provides the fastest response time in the readings, but less rejection of
noise.
3.4.6.7.9
PONS
– Remote Inhibit
Configures the power-ON remote inhibit mode either Latching or Live Mode for
selected channel; refer Figure 3-75. Refer to section 3.4.6.2. The remote inhibit modes
are explained in detail.
Figure 3-75: PONS
– Remote Inhibit channel 1
Figure 3-76: PONS
– Remote Inhibit channel 2