Asterion DC Multioutput ASA Series Operational Manual
M330516-01
3-63
Figure 3-72: PONS
– Configure Full Scale Value Channel-2
Figure 3-73: PONS
– Configure Full Scale Value Channel-3
Entry
Description
Full Scale
for Voltage:
To Set the power-on FSC value of analog voltage program for
Selected Channels, default FSC is 10V for Voltage reference source
and 10
kΩ for Resistive reference source. FSC (Full scale) value can
be programmed from 5V to 10V for voltage source as reference and
5
kΩ to 10kΩ for resistive reference source.
Full Scale
for Current:
To Set power-on FSC value of analog current program for Selected
Channels, default FSC is 10V for Voltage reference source and
10
kΩ for Resistive reference source. FSC (Full scale) value can be
programmed from 5V to 10V for voltage source as reference and 5
kΩ
to 10
kΩ for resistive reference source.
3.4.6.7.8
PONS - Measurements Setting
The PONS Measurement Settings screen sets the number of readings to be used for
averaging at power-on for voltage and current measurements. Refer to Figure 3-74.
Up-arrow will return to PONS menu screen; Refer Figure 3-54. Home button will return
to HOME MENU screen 2; Refer Figure 3-14.