CI Maintenance
To Switch from the CI Source to an EI XTR, SS, or Inert Source
234
5977B Series MSD Operating Manual
To Switch from the CI Source to an EI XTR,
SS, or Inert Source
Procedure
1
From the Tune and Vacuum Control view, vent the MSD. (See
on page 136.) The software will prompt you for the appropriate
actions.
2
Open the analyzer. (See
“To Open the Analyzer Chamber”
3
Remove interface tip seal, spring, and knurled tip seal retainer if switching to
an EI SS or EI Inert source.
4
Install the EI source. (See
“To Install the EI XTR, SS, or Inert Source”
5
If installing an EI XTR source, locate the brown extractor wire from storage
and connect it to the extractor lens and the source board.
6
Place the CI source in the ion source storage box.
7
If installing an EI SS or Inert source, remove the interface tip seal, spring, and
knurled tip seal retainer, and put them in the CI storage container.
8
Pumpdown the MSD. (See
CAUTION
Always wear clean gloves while touching the analyzer or any other parts that
go inside the analyzer chamber.
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap, and take other antistatic precautions
before
you open the
analyzer chamber. (See
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......