CI Maintenance
To Assemble the CI Source
244
5977B Series MSD Operating Manual
To Assemble the CI Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
Procedure
Refer to the exploded parts view
and the CI source parts list
on page 245 while using this procedure.
1
Assemble the ion focus lens, entrance lens, and lens insulator.
2
Slide the drawout plate and the drawout cylinder into the source body.
3
Slide the parts assembled in step 1 into the source body.
4
Install the setscrew that holds the lenses in place.
5
Attach the ceramic disk to the repeller and place on top of the source body.
6
Place the heater block assembly on top of the source body.
7
Reinstall the dummy filament, and the filament and attach with the
setscrews.
CAUTION
Always wear clean gloves when working in the analyzer chamber to avoid
contamination.