5
Operating in EI Mode
To Set Analyzer Temperatures from the Instrument Control View
122
5977B Series MSD Operating Manual
To Set Analyzer Temperatures from the
Instrument Control View
MSD ion source and Quad temperatures are stored in the current tune (*.u) file.
When a method is loaded, the setpoints in the tune file associated with that
method are downloaded automatically.
Procedure
1
In
Instrument Control
view, select
MS Temperatures
from the
Instrument
menu. (See
.)
2
Enter the
MS Source
and
MS Quad
(mass filter) temperatures in the
Setpoint
and
Limit
fields. (See
.)
The GC/MSD interface, ion source, and quadrupole heated zones interact.
The analyzer heaters may not be able to accurately control temperatures if
the setpoint for one zone is much different from that of an adjacent zone.
Figure 37 MS temperatures dialog
Table 9
Recommended temperature settings
EI operation
PCI operation
NCI operation
MS Source
230
250
150
MS Quad
150
150
150
CAUTION
Do not exceed 200 °C for the quadrupole or 350 °C for the source.