8
CI Maintenance
To Connect/Disconnect EI HES Model Wiring to the CI Source
5977B Series MSD Operating Manual
237
To Connect/Disconnect EI HES Model Wiring
to the CI Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Pliers, long-nose (8710-1094)
•
Tweezers (8710-2460)
Procedure
1
Use pliers to connect/disconnect the ceramic board wire leads (red, white,
black, and gray) at the source connectors. (See
.)
2
Use pliers to connect/disconnect the source heater wire leads at the ceramic
source board. (See
Figure 77 Wiring between the ceramic source board and the source
Filament 1
Repeller
Dummy filament
Entrance lens
Ion focus
Source heater
Source RTD
Ceramic source board
Filament 1
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......