5
Operating in EI Mode
To Perform High-Mass Testing
132
5977B Series MSD Operating Manual
To Perform High-Mass Testing
Materials needed
•
PFHT High Mass checkout sample 3 × 1 mL (5188-5357)
Procedure
1
Load the autotune file (atune.u, HES_atune.u, etune.u) then auto tune the
MSD. (See
2
Resolve the PFHT.M method under
MassHunter\MSD\x\methods\checkout\PFHT.M
where x is the instrument number being used.
3
Update and save the method.
4
Load the PFHT calibration sample into a vial and place in position 2.
5
In
Instrument Control
view select
High Mass Check
from the
Checkout
menu.
6
Follow the instructions on screen.
7
The run is completed, and results are printed within 5 minutes. (See