5
Operating in EI Mode
To Tune the MSD in EI Mode
5977B Series MSD Operating Manual
127
To Tune the MSD in EI Mode
You can also use the 7890B GC control panel to run the autotune that is currently
loaded in MassHunter. (See
“Operating the MSD from the 7890 GC Control
Procedure
1
Load the method that will be used for data acquisition.
2
In the
Instrument Control
view, verify the correct tune file is displayed in the
tile bar. For most applications,
Autotune
gives good results.
3
To select a different tune file select
MS Tune File
from the
Instrument
menu
to display the
Select Tune File
dialog. The
Settings
area displays the
important parameters for a selected tune file.
The tune file must match the type of ion source in the analyzer. If you are
using an EI source, select a tune file created for an EI source.
4
Click the
MS
Tune
icon
to display the
Select Tune Type
dialog. (See
.)
5
Select
Tune MSD
to perform a complete autotune, or select
Quick Tune
to
adjust peak width, mass assignment, and abundance, without changing ion
ratios.
6
Click
OK
to close this dialog and start the tune. If the MSD temperatures are
not stable, you are prompted to wait or override the wait by clicking
Override
.
Figure 40 Select tune type dialog
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......