5
Operating in EI Mode
To Set the JetClean Parameters for the Clean Only Mode
130
5977B Series MSD Operating Manual
To Set the JetClean Parameters for the Clean
Only Mode
Procedure:
1
In
Instrument Control
view, click the
MS Parameters
icon.
2
In the Single Quadrupole MS Method Editor, select the
JetClean
tab. (See
.)
3
From the
Operation
dropdown menu, select
Clean Only
.
4
Enter the parameters supplied by your method developer.
5
Save the method.
Figure 42 JetClean set to Clean Only mode