6
Operating in CI Mode
To Operate the Reagent Gas Flow Control Module
5977B Series MSD Operating Manual
151
To Operate the Reagent Gas Flow Control
Module
Procedure
1
In the
Manual Tune
dialog, click the
CI
tab to access the parameter settings
for controlling the CI gas flow. (See
CAUTION
After the system has been switched from EI to CI mode, or vented for any other
reason, the MSD must be baked out for at least 2 hours before tuning.
CAUTION
Continuing with CI autotune if the MSD has an air leak or large amounts of
water will result in
severe
ion source contamination. If this happens, you need
to
vent the MSD
and
clean the ion source
.
Figure 49 Manual tune CI tab