6
Operating in CI Mode
To Monitor CI Mode High Vacuum Pressure
5977B Series MSD Operating Manual
165
To Monitor CI Mode High Vacuum Pressure
Procedure
1
Start up and pumpdown the MSD. (See
2
In the
Tune and Vacuum Control
view select
Turn Vacuum Gauge on/off
from the
Vacuum
menu.
3
In the
Instrument Control
view you can set up an MS Monitor for reading. The
vacuum can also be read on the LCP or from the Manual Tune screen.
The gauge controller will not turn on if the pressure in the MSD is above
approximately 8 × 10
-3
Torr. The gauge controller is calibrated for nitrogen, but all
pressures listed in this manual are for helium.
The largest influence on operating pressure is the carrier gas (column) flow.
Typical pressures for various helium carrier gas flows are approximate and vary
from instrument to instrument. (See
Typical pressure readings
Use the G3397B Micro-Ion vacuum gauge. Note that the MFC is calibrated for
methane and the vacuum gauge is calibrated for nitrogen, so these
measurements are not accurate, but are intended as a guide to typical observed
readings. (See
on page 166.) They were taken with the following set of
conditions. Note that these are typical PCI temperatures:
WARNING
If hydrogen is used as a carrier gas or JetClean system supply, the carrier and
JetClean system supply shutoff valves must be closed before turning off the
MSD power. If the foreline pump is off, hydrogen will accumulate in the MSD
and an explosion may occur. Read
on page 24 before
operating the MSD with hydrogen gas.
Source temperature
250 °C
Quad temperature
150 °C
GC/MSD Interface
temperature
280 °C
Helium carrier gas flow
1 mL/min
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......