5
Operating in EI Mode
To change the network settings of the MSD
110
5977B Series MSD Operating Manual
6
Confirm that
BootP
is set to
OFF
. If your LAN assigns IP addresses using a
BootP server, click
ON
and skip the next step.
7
To update the
MSD IP address
,
Gateway IPA
, and
SubNet Mask
enter the
new values. Before clicking submit, you can recover the previous settings by
clicking
Return to Main Menu
and then returning here.
8
Click
Submit
to upload this net configuration to the MSD.
A dialog opens to confirm that the network configuration process has
started. (See
.)
9
Click
OK
to close the dialog and wait for the prompt requesting to
Manually
reboot of the MSD/SmartCard to activate the new Settings
.
10
Use the
MSD on/off
switch to reboot the MSD SmartCard.
Figure 30 IP address configuration