To Replace an Intuvo 9000 GC Gasket
This procedure assumes you have already removed the column, 9000 GC/MS
Tail, or other part that
sits on top of the gasket and that the instrument
components are below 40
°C.
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Tweezers (8710-2460)
•
Intuvo gasket, polyimide 5/pk, for temperatures < 350 °C (5190-9072)
•
Intuvo gasket, nickel 5/pk, for temperatures 350 °C to 450 °C (5190-9073)
Procedure
1
Remove the gasket tab from the alignment stud and discard the old gasket.
Tweezers help when you are wearing the required gloved.
2
If needed, install any inlet or detector chips. All chips must be installed before
installing the new gasket.
3
Carefully remove the new gasket from its packaging. Inspect the gasket to be
sure it is not deformed. The two round lobes are the sealing surfaces.
4
Carefully insert the round gasket lobes into the bus click and run fitting. Note
that the gasket is double-sided.
5
Place the hole in the gasket over the alignment stud in the bus fitting, and
press the gasket body flat against the bus.
6
Check that the gasket's round lobes fit flat against the bus click and run fitting.
The new gasket is ready for the chip or column attachment.
WARNING
The inlet, detector, bus components, and column can be hot enough to cause
burns. Cool heated zones to a safe handling temperature before continuing.
CAUTION
Always wear clean gloves while handling any parts that are in the flow path
going to the analyzer or parts that are located inside the analyzer.
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......