2
Installing 8890 GC Columns
To Install the GC/MSD Interface Tip Seal
50
5977B Series MSD Operating Manual
To Install the GC/MSD Interface Tip Seal
Materials needed
•
Interface tip seal (G3870-20542)
•
Tip seal spring (G7005-20024)
•
Knurled tip seal retainer (G3870-20547)
The interface tip seal must be in place for the CI source, the EI XTR source,
and HES. The interface tip seal, spring, and knurled seal retainer are not
used with EI SS or Inert source.
•
Gloves, clean, lint-free
•
Large 8650-0030
•
Small 8650-0029
Procedure
1
Verify that the CI source, EI XTR source, or HES is installed. This tip seal and
its spring must not be installed when an EI SS or Inert source is installed (See
2
Remove the interface tip seal, spring, and knurled nut retainer from the ion
source storage box. In this order, slide the spring, tip seal, and knurled tip seal
retainer over the column sleeve.
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap and take other antistatic precautions
before
you open the
analyzer chamber.
CAUTION
Always wear clean gloves while handling any parts that go inside the GC or the
analyzer chambers.
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......