4
Installing 7890 GC Columns
To Calibrate Column Flow Linear Velocity in a 7890 GC
5977B Series MSD Operating Manual
95
To Calibrate Column Flow Linear Velocity in a
7890 GC
The steps listed below apply to 7890 and 6890 series GCs only. Capillary
columns must be calibrated prior to use with the MSD.
Procedure
1
Set Data Acquisition for splitless manual injection, and set up a real time plot
to monitor
m/z
28.
2
Press [
Prep Run]
on the GC keypad.
3
Inject 1 µL of air into the GC inlet and press [
Start Run
].
4
Wait until a peak elutes at
m/z
28. Note the retention time.
5
In the
Instrument Control
view, select
GC Parameters
from the
Instrument
menu.
6
Select the
Configuration
tab and then select the
Columns
tab.
7
Select your installed column from the table.
8
Click
Calibrate
to display the
Calibrate Column
dialog.
9
Click
Calc Length
in the
If unretained peak holdup time is known
section to
display the
Calculate Column Length
dialog. (See