4
Installing 7890 GC Columns
To Calibrate Column Flow Linear Velocity in a 7890 GC
96
5977B Series MSD Operating Manual
10
Verify that the parameters listed (temperature, inlet and outlet pressures, and
gas type) are those used in the method to determine the holdup time.
Change any parameters that are different than those used in your method.
11
Enter the recorded retention time in the
Holdup Time
field. Move the cursor to
another parameter’s field and the calibrated column length appears.
12
Click
OK
to save the changes and exit the dialog.
13
Click
OK
on the
Calibrate Columns
dialog to save the calibration.
With capillary columns, such as those used with the MSD, linear velocity is often
measured rather than volumetric flow rate.
Figure 25.
Calibrate column length dialog