6
Operating in CI Mode
To Verify NCI Performance
164
5977B Series MSD Operating Manual
To Verify NCI Performance
This procedure is for EI/PCI/NCI MSDs
only.
Materials needed
•
OFN, 100 fg/µL 3 × 1 mL (5188-5347)
Procedure
1
Verify that the MSD performs correctly in EI mode.
2
Load the
NCICH4.U
tune file, and accept the temperature setpoints.
3
Select
Gas A
and set flow
to 40%.
4
In
Tune and Vacuum Control
Autotune (Methane Reagent Gas)”
Note that there are no criteria for a “passing” Autotune in CI. If the Autotune
completes, it passes.
5
Run the NCI sensitivity method: OFN_NCI.M using 2 µL of 100 fg/µL OFN.
6
Verify that the system conforms to the published sensitivity specification.
Please refer to the Agilent Web site at
for
specifications.
CAUTION
Always verify MSD performance in EI before switching to CI operation. (See
“To Verify EI System Performance”
on page 131.) Always set up the CI MSD in
PCI first, even if you are going to run NCI.
Summary of Contents for 5977B CI MSD
Page 1: ...Agilent 5977B Series MSD Operating Manual ...
Page 12: ...12 5977B Series MSD Operating Manual ...
Page 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Page 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Page 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Page 253: ......