Operating in CI Mode
To Pumpdown the MSD in CI Mode
148
5977B Series MSD Operating Manual
To Pumpdown the MSD in CI Mode
This procedure assumes that the instrument will eventually be PCI tuned using
methane after the system is stable.
Procedure
1
Follow the instructions for the EI MSD. (
After the software prompts you to turn on the GC/MSD interface heater and
GC oven, perform the following steps.
2
In the
Manual Tune
dialog, click the
Values
tab to monitor that the pressure is
decreasing (Hi-Vac gauge option installed).
3
In the
Manual Tune
dialog, click the
CI Gas
tab, then in the
Valve Settings
area close
Gas Valve A
,
Gas Valve B
, and the
ShutOff Valve
.
4
Verify that
PCICH4.U
is loaded (top left in
Manual Tune
dialog) and click the
Values
tab to accept the temperature setpoints.
Always start up and verify system performance in PCI mode before
switching to NCI.
5
Set the GC/MSD interface to 280 °C. (See
6
Set
Gas A (methane)
to 20%.
7
Let the system bake out and purge for at least 2 hours. If you will be running
NCI, for best sensitivity, bake out the MSD overnight.