To Install the EI XTR, SS, or Inert Source
5977B Series MSD Operating Manual
217
To Install the EI XTR, SS, or Inert Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Pliers, long-nose (8710-1094)
Procedure
1
Slide the ion source into the source radiator. (See
2
Insert and hand tighten the source thumbscrews. Do not overtighten the
thumbscrews. (See
.)
3
Connect the ion source wires. (See
“To Connect/Disconnect Wiring from the
EI XTR, SS, and Inert Sources”
4
Close the analyzer chamber. (See
“To Close the Analyzer Chamber”
Figure 70.
Installing the EI source
Ceramic source board
Ion source
Thumbscrews
Source heater and
temperature sensor wires
Source radiator
Extractor wire