ZEISS
3 Product and Functional Description | 3.3 Optional Components and Accessories
7
Wobble
Sweeps the acceleration voltage. If the aperture is slightly misaligned, a shift in X and/or
Y direction can be observed.
8
Freeze
Stops the scan and grabs one complete frame at the current imaging conditions.
9
Exchange | Resume
§
Exchange
Starts the pre-defined macro for specimen exchange with the airlock.
§
Resume
Starts the pre-defined macro to finish specimen exchange with the airlock.
10
Camera
Switches to chamber view.
11
Focus | Scan Speed +/−
§
Focus
Changes the focal point of the column by adjusting the magnitude of the objective
lens.
§
Scan Speed +/−
Increases (+) or decreases (−) the scan speed by doubling or halving the beam dwell
time with each click step.
Instruction Manual ZEISS Crossbeam 350 | en-US | Rev. 3 | 349500-8111-000
51