8.2 Error Messages
Error Messages
Messages may appear on the screen during operation. This section describes the error messages and
how to respond to them. Messages are displayed according to the language setting (see section 7.6).
If the corrective action states that serving is required, contact your nearest YOKOGAWA dealer.
In addition to the following messages, there are also communications error messages. These error
messages are described in the Communication Interface User’s Manual (IM AQ7280-17EN).
Error in Execution
Code
Message
20
Network Option is not installed.
21
Power Monitor Option is not installed.
22
Light source option is not installed.
24
One or more conditions in this file are not supported by this product.
27
The end point is different between acquired data and reference data. Please check the connected fiber cable.
28
The measurement may not be completed within the specified duration.
29
In remote control mode, all keys are locked except F1 key. Please hit F1 key to exit the remote control mode.
30
In Local Lockout mode, all keys are locked. Please cancel the local lockout.
33
File operation is cancelled.
35
The optical plug may not be connected securely.
36
Either of the analysis information below could not be retrieved. (Cursor info, Free marker info., Event info.)
37
The end point distance does not match with the master end point.
40
No more averaging is allowed. (already MAX averaging condition) Set "Average Continue"OFF for new
measurement.
41
Now updating firmware. Please wait. It takes approx. 20 seconds. Do not turn off power during update.
Restart automatically after successful updating.
46
A lot of files are saved in the root folder. The file manipulation slows when saved any further. Please make a folder
in the root folder, and save files in that.
47
Please start measurement after executing unchecking this fiber.
48
Now Logging. Please Wait.
49
ESC: Close the Touch Screen Test window, ENTER: Clear the display
60
Now Logging. Please Wait.
61
Logging process is completed. Please Save File.
62
Now adjusting LS power. Please wait.
63
LS power adjustment is completed.
64
Failed to adjust LS power. Please check whether the patch fiber cable is properly connected between LS port and
OPM port.
69
Optical fiber is found at the cursor position. Please stop the measure by pressing "REALTIME".
500
The measurement condition is failed. Please reset ****** in SETUP menu.
501
Not executable during measurement. Please stop the measurement and execute again.
502
The measurement cannot be started during printing. Please either wait until the end of the printing or interrupt the
printing.
503
The measurement cannot be started during the file operation. Please either interrupt the file operation or wait until
the end of the file operation.
504
The measurement could not be completed within the specified averaging duration. Please modify the averaging
duration.
505
Hardware failed, and needs to be repaired. Please contact Yokogawa's representatives.
506
Hardware failed, and needs to be repaired. Please contact Yokogawa's representatives.
507
Hardware failed, and needs to be repaired. Please contact Yokogawa's representatives.
508
Hardware failed, and needs to be repaired. Please contact Yokogawa's representatives.
509
Hardware failed, and needs to be repaired. Please contact Yokogawa's representatives.
510
PLUG CHECK Error. Please check or clean the connector.
511
This parameter is not valid in Simple (Full Auto) mode.
512
ZERO SET ERROR