2-24
IM AQ7280-01EN
2.
Using the
rotary knob
and
ENTER
, select the route to be measured. The Select Route screen
closes.
Fault Locator Screen
3.
Press
AVG
. The measurement begins, and the measurement waveform is displayed on the
screen. When the averaged measurement completes, the distance to a break is automatically
displayed.
Set the distance
reference.
See section 4.1.
OTDR function
Switches to the screen that
appears when OTDR is
selected from the top
menu. You cannot return to
the original screen.
Set the wavelength.
See page 2-2.
Distance to a break
Selected route
Waveform display screen
See section 3.1.
Cursor position from the distance reference
See section 4.1.
Note
You can save measured data, but saved waveform data cannot be loaded on the AQ7280. For details on
saving measured data, see section 6.3.
Changing Settings
Press
SETUP
to display the following screen.
Set the end of fiber.
See page 2-11.
Change the selected route.
The Select Route menu will appear. See step 2.
2.4 Performing Averaged Measurement and Measuring the Distances to Breaks