2-39
IM AQ7280-01EN
Optical Pulse Measurement (OTDR)
1
2
3
4
5
6
7
8
Index
Averaged Measurement
3.
Press
AVE
to start an averaged measurement.
The screen shows the measurement waveform, and the Open Trace menu appears. During measurement,
a mark appears at the top of the display to indicate that the laser light is on.
Set markers.
See section 4.1.
Cannot be selected
during measurement.
Cannot be selected
during measurement.
Create labels.
You can enter a label for each
core. Enter the label characters
by following the procedure
provided in “Entering Character
Strings” in the IM AQ7280-02EN.
Note
• If Wavelength on the main view screen is set to “All,” measurement is executed using multiple
wavelengths for each core.
• If you enter the label characters in the label setting of the project setup screen, the same label characters
can be entered at once for all cores.
When Auto Save is Set to ON in the Project Measurement Conditions
When measurement is complete, the measured results are automatically saved. The main view
screen appears, and pass/fail judgment indications appear in the core number cells whose
measurement results have been saved.
When Auto Save is Set to OFF in the Project Measurement Conditions
When measurement is complete, the measured results are not automatically saved. All the items on
the Open Trace menu become available. When returning to the main view screen, a message for
confirming the saving of the measured results appears. If necessary, you can save the measured
results.
2.8 Performing Multi-Core Fiber Measurement