5-10
IM AQ7280-01EN
• Auto Zoom
The vertical zoom factor is set automatically so that all the logging data is displayed.
The median value between the maximum and minimum of the logging data is the center position
on the screen.
If there is no logging data, auto zoom is not performed.
• Zoom Initialize
Zoom initialization returns the vertical and horizontal zoom factors to ×1.
Note
The logging results are saved in a CSV file, so you can open it with a spreadsheet application. The file can
also be loaded into the AQ7280 to display the graphs.
Setting the Logging Conditions
3.
Press
SETUP
and then the
Logging
soft key to display the following screen.
Set the measurement interval
(500ms, 1s, 2s, 5s, 10s).
Set the number of times
(10 to 36000).
Logging time
The time to perform logging measurements over is calculated automatically
from the measurement interval and the logging count and displayed.
Interval
Set the interval for logging the measured values from the following.
500ms, 1s, 2s, 5s, 10s
Times
Set the number of times to log in the following range.
10 to 36000
5.2 Measuring Optical Power (OPM, /PC option)