4-9
IM AQ7280-01EN
Optical
Pulse Analysis
1
2
3
4
5
6
7
8
Index
Measuring Splice Loss
When Marker Mode Is Set to Marker
1.
From the top menu, select
OTDR
.
2.
Press the
Marker
soft key and then the
4 Point Markers
soft key to display the following
screen.
Measurement start position
Set marker .
1
2
Splice loss event start position
Set marker .
Splice loss event end position
Set marker Y2.
Measurement end position
Set marker .
3
Measured splice loss (result)
When Marker Mode Is Set to Line
1.
From the top menu, select
OTDR
.
2.
Press the
Marker
soft key and then the
5 Point Markers
soft key to display the following
screen.
Near-end point
Set marker n.
Approximated line point on
the near-end side
Set marker N.
Splice loss detection point
Set marker E.
Approximated line point on
the far-end side
Set marker F.
Far-end point
Set marker f.
Measured splice loss (result)
4.1 Analyzing Waveforms