2-48
IM AQ7280-01EN
Main View Screen
1.
From the top menu, select
Schedule Measurement
to display the following screen.
Select the displayed
information (Logging, Trace).
Switch the trace on
which to place the
cursor (Trace1 to 4).
Set the auxiliary function.
Set the display items.
Starts or stops schedule
measurement
Measurement
end time
Date and time at the cursor
position and measurement
item value
Name of the file currently being saved to
Schedule-measurement-in-progress indication
Time remaining until the next optical pulse measurement
Overview display
When the displayed information is logging, the waveform at the cursor position
is displayed. When the displayed information is trace, logging data is displayed.
Scale value
Displayed Information
• Logging
The measured optical pulse power values are logged on the time axis. You can view the places
where the optical power is changing. Also, you can view the date and time and the power value
at the cursor position.
• Trace
You can display the waveform data at the cursor position. While waveform data is being
displayed, the menu changes.
Select the displayed waveform.
Switches to the waveform display of each wavelength.
Event analysis
See section 4.2.
Marker
See section 4.1.
Starts or stops schedule
measurement
2.9 Monitoring Optical Fiber Cables (/MNT option)