8-3
IM AQ7280-01EN
Troubleshooting, Maintenance, and Inspection
1
2
3
4
5
6
7
8
Index
Code
Message
513
ZERO SET ERROR
514
Exceeding limit. It may cause damage of the instrument. Please disconnect the plug.
515
ZERO SET ERROR
516
Fiber In Use Alarm Error 1
517
Fiber In Use Alarm Error 2
518
ZERO SET ERROR
519
ZERO SET ERROR
520
Hardware failed, and needs to be repaired. Please contact Yokogawa's representatives.
521
Operation is stopped outside of operating environmental condisions
522
Hardware failed, and needs to be repaired. Please contact Yokogawa's representatives.
524
Failure to initialize the option module.
525
The Interval is too short. Please change the inteval value.
526
Please set Avg Duration.
527
Need the AC power supply to measure. Please connect the AC adapter.
528
Free space is not enough. Please set the interval or the period again.
550
The event list does not exist in either of trace or both.
551
Measurement range of those two traces are not same reference points. Please set the origin (0m).
552
Fiber lengths do not match. The end point distance error between two traces must be within 3%.
553
Total number of events exceeds 100 for the 2-way analysis.
554
2-Way Trace cannot be executed due to difference in wavelengths of those two traces.
555
The differential trace cannot be created. Trace conditions are not same.
556
The differential trace cannot be created. No trace data available.
557
The differential trace cannot be created. The sampling resolutions are not same.
558
The differential trace cannot be created. The group indices are not same.
559
The differential trace cannot be created. The start points are not same.
560
When the event fix is ON, the distance reference setup cannot be performed.
561
1
-
2
has not been set. This function is effective only when
1
-
2
is set.
562
n-E has not been set. This function is effective only when n-E is set.
563
Conditions of the macro bending do not match.
601
Cannot save onto the media. This media is for read-only.
602
Can not recognize file system. -Internal memory: Contact Yokogawa's representatives. - USB memory or SD
card: Try the other media or format again with FAT
604
Storage media has been disconnected while the media is being accessed.
605
Same file name or folder name exists.
608
Invalid file name or folder name
609
Invalid file format
612
Invalid path name
614
Unknown file or folder
616
Cannot save onto the media. This media is for read-only.
620
Free space is not enough.
622
Unknown folder
623
Folder is not empty.
629
Writing to USB memory or SD card is not allowed.
634
Failed to delete the folder. Please confirm the following. - Media is correctly installed. - Media is formatted. -
Media is not read-only.
635
Failed to delete the folder. The path name or the file name is too long.
636
Failed to delete the folder. The hierarchy below the specified folder is too deep.
637
Failed to delete the folder. The path name is invalid.
638
The folder could not be deleted. Failed to delete files in the folder. Please confirm the following. - Media is correctly
installed. - Media is formatted. - Media is not read-only.
639
Failed to copy the folder. Please confirm the following. - Media is correctly installed. - Media is formatted. - Media
is not read-only.
640
Folder cannot be copied. The path name or the file name is too long.
641
Failed to copy the folder. The hierarchy below the specified folder is too deep.
642
Failed to copy the folder. The path name is invalid.
643
The folder copy could not be completed. Failed to create a destination folder. Please confirm the following.
- Media is correctly installed. - Media is formatted. - Media is not read-only.
644
The folder copy could not be completed. Failed to copy the files in the folder. Please confirm the following.
- Media is correctly installed. - Media is formatted. - Media is not read-only.
645
The path name is too long.
646
USB memory or SD card can not be recognized.
8.2 Error Messages