8-5
IM AQ7280-01EN
Troubleshooting, Maintenance, and Inspection
1
2
3
4
5
6
7
8
Index
Code
Message
803
FTP Error : Rm Please confirm network connection and setting.
804
FTP Error : List Please confirm network connection and setting.
805
FTP Error : Mkdir Please confirm network connection and setting.
806
FTP Error : Rmdir Please confirm network connection and setting.
807
FTP Error : Get Please confirm network connection and setting.
808
FTP Error : Put Please confirm network connection and setting.
809
FTP Error : GetData Please confirm network connection and setting.
810
FTP Error : PutData Please confirm network connection and setting, and desk capacity.
811
FTP Error : AppedData Please confirm network connection and setting, and desk capacity.
812
FTP Error : Client Handle Please confirm network connection and setting.
813
FTP Error : Others Please confirm network connection and setting.
814
Failed network initialize. Please confirm network connection and setting.
815
Different passwords are entered. Please try again.
817
Test Error occurred.
818
Test succeeded.
819
ROM Test Error occurred.
820
RAM Test Error occurred.
821
System File Test Error occurred.
822
Battery Test Error occurred.
823
Temperature Test Error occurred.
854
In USB Storage mode, all keys are locked. Please disconnect the USB Cable.
855
While a measurement is progressed or a file is accessed, it cannot go to USB storage mode. Please try again after
stopping a measurement or a file access.
900
Backup data is damaged. this instrument starts up with the factory setting.
901
Optical connector is not correctly connected. Please clean and reconnect the connector again.
902
Battery is low Please power it off, and charge the battery or replace the battery. Or, please use the AC power
supply.
903
Backup battery failed, and needs to be repaired. Please contact Yokogawa's representatives.
904
Hardware failed, and needs to be repaired. Please contact Yokogawa's representatives.
905
Battery error. Please replace the battery.
906
Battery is low The instrument will be powered off in 10 sec.
907
Battery temperature is too high. The instrument will be powered off in 10 sec. Please turn off power to cool battery
908
Battery temperature is too low. The instrument may be damaged if used in this condition. The instrument will be
powered off in 10 sec. Please turn off power and use battery with proper operating temperature.
909
The temperature inside the instrument is too high. The instrument may be damaged if used in this condition. The
instrument will be powered off in 10 sec. Please turn off power to allow instrument to cool down.
910
The temperature inside the instrument is too low. The instrument may be damaged if it is used in this condition.
The instrument will be powered off in 10 sec. Please turn off power until internal temperature is increased.
911
The Voltage of AC power supply is too low. The instrument may be damaged if it is kept using in this condition.
The instrument will be powered off in 10 sec. Please make sure of using the dedicated AC adapter.
912
The Voltage of AC power supply is too high. The instrument may be damaged if it is kept using in this condition.
The instrument will be powered off in 10 sec. Please make sure of using the dedicated AC adapter.
913
Battery is low. The instrument may be damaged if it is kept using in this condition. The instrument will be powered
off in 10 sec. Please use AC adapter.
914
Errors in the battery or in the charging circuit. The instrument may be damaged if it is kept using in this condition.
The instrument will be powered off in 10 sec. Please remove the battery and use AC adapter.
915
Battery temperature is too high. Please stop measuring, and wait for a while.
916
Battery temperature is too high. The measurement was aborted. Please power it off.
917
The temperature inside the instrument is increasing. Please stop measuring, and wait for a while.
918
The temperature inside the instrument is too high. The measurement was aborted. Please power it off.
919
Non-standard battery is connected. The instrument may be damaged if it is kept using in this condition. The
instrument will be powered off in 10 sec. Please replace the battery.
921
There is incoming light. Or, optical module may be damaged.
922
Incorrect date and time setting. Set the correct date and time.
923
The ambient temperature is out of range. The light source cannot be turned on.
925
Please use AC adapter.
8.2 Error Messages