3-1
IM AQ7280-01EN
Optical Pulse W
aveform Display
1
2
3
4
5
6
7
8
Index
3.1 Waveform Display Screen
Display Screen
Date and time
See the Started Guide, IM AQ7280-02EN.
Value per division on the vertical axis
The OTDR port emitting light
Waveform display area
Measurement mode
See section 2.1.
Progress during averaged measurements (AVG)
The progress bar behind the characters indicates progress.
See section 2.2.
Cursor movement resolution
See section 3.3.
Battery level/AC
See the Started Guide, IM AQ7280-02EN.
File name
The name of the measurement
data file that is currently displayed
For details on file names, see
section 6.5.
Position of the left edge of the screen
When the distance reference is set, the position of
the distance reference is 0, and this displays the
distance from the distance reference to the position
of the left edge of the screen as a negative value.
Measurement conditions
See section 2.1.
Sample interval
See page 2-6.
Distance per division on the horizontal axis
Position of the right edge of the
screen
When the distance reference is
set, this displays the distance
from the distance reference to the
position of the right edge of the
screen.
Distance from the horizontal axis’
measurement reference point to
the cursor position
See section 4.1.
Full screen display of the waveform display
area
The section that is displayed in the waveform
display area is represented by a box (overview).
See section 3.3.
Marker section measured values
The attenuation, distance, and attenuation
per km between the markers are displayed.
Shortcut display
Shortcut feature for starting measurement, saving
data, and so on.
Marker approximation method
See pages 2-3 to 2-4.
Shortcut Display
Tapping a shortcut icon performs one of the following functions.
Icon
Function
Preset marker
A
is applied to the waveform. (see section 4.1).
Preset marker
B
is applied to the waveform. (see section 4.1).
Starts or stops real-time measurement (see section 2.5). The icon toggles every time you tap.
Starts or stops averaged measurement (see section 2.2). The icon toggles every time you tap.
Initializes the scale (see section 3.3).
*
Makes a reference trace (see section 2.7).
Displays a file list (see section 6.3).
*
Converts waveform data into a PDF file (see section 6.8).
Saves waveform data (see section 6.3).
* These are invalid in multi-core fiber measurement.
Chapter 3
Optical Pulse Waveform Display