2-12
IM AQ7280-01EN
Advanced Analysis
• Pass Fail Judgement
Using the
rotary knob
and
ENTER
, set Pass Fail Judgement to
Setup[OFF]
or
Setup[ON]
. The
following screen appears.
Set the judgment (Hide, Display).
Set the splice loss threshold
(0.01dB to 9.99dB).
Set the return loss threshold
(20dB to 70dB).
Set the splitter loss threshold
(1.0dB to 20.0dB).
Set the dB/km loss threshold
(0.01dB to 9.99dB).
Set the total loss
threshold (1dB to 65dB).
Set the preset (UserDefine,
Preset 1, Preset 2, Preset 3).
Turns the pass fail judgment on
or off.
Set the connector loss
threshold (0.01dB to 9.99dB).
Pass Fail Judgement
A judgment is performed on the events detected in the target waveform, and those that exceed
the specified threshold are displayed as fault events.
ON: Events that have exceeded their thresholds are displayed with an asterisk attached to
them.
OFF: Events that have exceeded their thresholds are not displayed with an asterisk attached
to them.
Judgment
The judgment information is displayed in the lower left of the AQ7280 screen.
Display: The number of fault events, total loss, and total return loss are displayed.
Hide: The judgment information is not displayed.
Events judged as Fail
Judgment display
2.1 Setting the Measurement and Analysis Conditions